a) Check that all the analytical balance's glass doors are closed.
b) Ensure that the analytical balance is balanced.
Just firmly press, DO NOT SMASH micropipette into the tip.
CRITICALThis allows the air volume's humidity to stabilize.
- Download the template of the record tables: MICROPIPETTE_CALIBRATION_TBLS.pdf
- record the weight on a table.
- Press Zero (Tare) after each measurement to reset the balance back to zero.
- Make notes for each record table in the “observations” pane of the attached tables.
Benchling is actively tested against the latest versions of Chrome, Firefox, Safari, and Edge. It doesn't look like your current browser is supported - for more information, click here.
Evaluate micropipette calibration status by constructing a volume table and determining mean pipette volume, standard deviation and coefficient of variation. Calculations can be done with Excel or R.